| 48 T | $320.00 |
| 96 T | $458.00 |
| 96 T*5 | $1947.00 |
| Concentration (pg/mL) | OD | Corrected OD |
|---|---|---|
| 500.00 | 2.105 | 2.006 |
| 250.00 | 1.630 | 1.531 |
| 125.00 | 1.158 | 1.059 |
| 62.50 | 0.860 | 0.761 |
| 31.25 | 0.640 | 0.541 |
| 15.63 | 0.359 | 0.260 |
| 7.81 | 0.154 | 0.055 |
| 0.00 | 0.099 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 84-97% | 91% |
| EDTA plasma(n=5) | 88-114% | 101% |
| Heparin plasma(n=5) | 80-93% | 87% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 94-117% | 90-113% | 82-100% | 83-96% |
| EDTA plasma(n=5) | 88-110% | 84-100% | 85-95% | 88-97% |
| Heparin plasma(n=5) | 84-90% | 93-114% | 85-108% | 89-106% |