| 48 T | $320.00 |
| 96 T | $458.00 |
| 96 T*5 | $1947.00 |
| Concentration (pg/mL) | OD | Corrected OD |
|---|---|---|
| 5000.00 | 2.157 | 2.076 |
| 2500.00 | 1.602 | 1.521 |
| 1250.00 | 1.185 | 1.104 |
| 625.00 | 0.831 | 0.750 |
| 312.50 | 0.515 | 0.434 |
| 156.25 | 0.252 | 0.171 |
| 78.13 | 0.191 | 0.110 |
| 0.00 | 0.081 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 77-95% | 86% |
| EDTA plasma(n=5) | 92-106% | 99% |
| Heparin plasma(n=5) | 88-103% | 95% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 87-95% | 85-97% | 89-101% | 85-97% |
| EDTA plasma(n=5) | 78-91% | 86-92% | 89-96% | 91-97% |
| Heparin plasma(n=5) | 95-104% | 97-105% | 82-95% | 89-97% |