| 48 T | $412.00 |
| 96 T | $588.00 |
| 96 T*5 | $2499.00 |
| Concentration (ng/mL) | OD | Corrected OD |
|---|---|---|
| 100.00 | 2.168 | 2.080 |
| 50.00 | 1.569 | 1.481 |
| 25.00 | 1.217 | 1.129 |
| 12.50 | 0.937 | 0.849 |
| 6.25 | 0.566 | 0.478 |
| 3.13 | 0.315 | 0.227 |
| 1.57 | 0.175 | 0.087 |
| 0.00 | 0.088 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 85-97% | 91% |
| EDTA plasma(n=5) | 80-94% | 87% |
| Heparin plasma(n=5) | 91-105% | 98% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 86-97% | 88-95% | 92-101% | 87-98% |
| EDTA plasma(n=5) | 91-99% | 85-93% | 82-95% | 93-99% |
| Heparin plasma(n=5) | 85-97% | 83-96% | 95-103% | 88-95% |