| 48 T | $320.00 |
| 96 T | $458.00 |
| 96 T*5 | $1947.00 |
| Concentration (pg/mL) | OD | Corrected OD |
|---|---|---|
| 4000.00 | 2.168 | 2.080 |
| 2000.00 | 1.569 | 1.481 |
| 1000.00 | 1.217 | 1.129 |
| 500.00 | 0.937 | 0.849 |
| 250.00 | 0.566 | 0.478 |
| 125.00 | 0.315 | 0.227 |
| 62.50 | 0.175 | 0.087 |
| 0.00 | 0.088 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 82-94% | 88% |
| EDTA plasma(n=5) | 97-107% | 102% |
| Heparin plasma(n=5) | 87-99% | 93% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 79-96% | 85-99% | 79-93% | 89-97% |
| EDTA plasma(n=5) | 88-103% | 86-92% | 91-99% | 95-102% |
| Heparin plasma(n=5) | 79-91% | 87-96% | 83-101% | 82-95% |