| 48 T | $320.00 |
| 96 T | $458.00 |
| 96 T*5 | $1947.00 |
| Concentration (pg/mL) | OD | Corrected OD |
|---|---|---|
| 5000.00 | 2.164 | 2.070 |
| 2500.00 | 1.394 | 1.300 |
| 1250.00 | 1.027 | 0.933 |
| 625.00 | 0.866 | 0.772 |
| 312.50 | 0.518 | 0.424 |
| 156.25 | 0.318 | 0.224 |
| 78.13 | 0.197 | 0.103 |
| 0.00 | 0.094 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 80-95% | 87% |
| EDTA plasma(n=5) | 90-105% | 97% |
| Heparin plasma(n=5) | 86-99% | 92% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 92-105% | 82-92% | 86-94% | 81-95% |
| EDTA plasma(n=5) | 91-97% | 86-92% | 79-93% | 89-106% |
| Heparin plasma(n=5) | 95-104% | 87-96% | 89-101% | 82-90% |