| 48 T | $412.00 |
| 96 T | $588.00 |
| 96 T*5 | $2499.00 |
| Concentration (ng/mL) | OD | Corrected OD |
|---|---|---|
| 20.00 | 2.082 | 1.987 |
| 10.00 | 1.709 | 1.614 |
| 5.00 | 0.980 | 0.885 |
| 2.50 | 0.797 | 0.702 |
| 1.25 | 0.519 | 0.424 |
| 0.63 | 0.218 | 0.123 |
| 0.32 | 0.241 | 0.146 |
| 0.00 | 0.095 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 83-95% | 89% |
| EDTA plasma(n=5) | 87-97% | 90% |
| Heparin plasma(n=5) | 81-95% | 88% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 80-101% | 87-96% | 96-105% | 95-103% |
| EDTA plasma(n=5) | 84-99% | 83-95% | 83-95% | 96-101% |
| Heparin plasma(n=5) | 85-100% | 84-98% | 97-101% | 88-96% |