| 48 T | $412.00 |
| 96 T | $588.00 |
| 96 T*5 | $2499.00 |
| Concentration (pg/mL) | OD | Corrected OD |
|---|---|---|
| 5000.00 | 2.166 | 2.057 |
| 2500.00 | 1.697 | 1.588 |
| 1250.00 | 1.167 | 1.058 |
| 625.00 | 0.899 | 0.790 |
| 312.50 | 0.536 | 0.427 |
| 156.25 | 0.461 | 0.352 |
| 78.13 | 0.114 | 0.005 |
| 0.00 | 0.109 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 88-102% | 95% |
| EDTA plasma(n=5) | 79-97% | 88% |
| Heparin plasma(n=5) | 83-95% | 89% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 89-97% | 85-94% | 79-96% | 85-99% |
| EDTA plasma(n=5) | 88-101% | 86-92% | 83-98% | 96-103% |
| Heparin plasma(n=5) | 85-94% | 87-103% | 81-95% | 82-90% |