| 48 T | $265.00 |
| 96 T | $376.00 |
| 96 T*5 | $1598.00 |
| Concentration (pg/mL) | OD | Corrected OD |
|---|---|---|
| 5000.00 | 2.231 | 2.149 |
| 2500.00 | 1.487 | 1.405 |
| 1250.00 | 1.069 | 0.987 |
| 625.00 | 0.874 | 0.792 |
| 312.50 | 0.479 | 0.397 |
| 156.25 | 0.384 | 0.302 |
| 78.13 | 0.248 | 0.166 |
| 0.00 | 0.082 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 98-104% | 102% |
| EDTA plasma(n=5) | 95-107% | 101% |
| Heparin plasma(n=5) | 96-104% | 100% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 86-95% | 93-103% | 87-98% | 85-96% |
| EDTA plasma(n=5) | 85-96% | 86-95% | 90-102% | 81-90% |
| Heparin plasma(n=5) | 87-95% | 83-96% | 92-102% | 82-93% |