| 48 T | $265.00 |
| 96 T | $376.00 |
| 96 T*5 | $1598.00 |
| Concentration (pg/mL) | OD | Corrected OD |
|---|---|---|
| 2000.00 | 2.068 | 1.973 |
| 1000.00 | 1.522 | 1.427 |
| 500.00 | 1.043 | 0.948 |
| 250.00 | 0.677 | 0.582 |
| 125.00 | 0.538 | 0.443 |
| 62.50 | 0.419 | 0.324 |
| 31.25 | 0.242 | 0.147 |
| 0.00 | 0.095 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 92-100% | 96% |
| EDTA plasma(n=5) | 90-121% | 106% |
| Heparin plasma(n=5) | 89-103% | 96% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 88-100% | 89-113% | 84-95% | 88-99% |
| EDTA plasma(n=5) | 92-113% | 81-109% | 89-99% | 86-97% |
| Heparin plasma(n=5) | 85-94% | 92-110% | 88-104% | 80-99% |