| 48 T | $265.00 |
| 96 T | $376.00 |
| 96 T*5 | $1598.00 |
| Concentration (pg/mL) | OD | Corrected OD |
|---|---|---|
| 2000.00 | 2.027 | 1.933 |
| 1000.00 | 1.528 | 1.434 |
| 500.00 | 1.043 | 0.949 |
| 250.00 | 0.661 | 0.567 |
| 125.00 | 0.539 | 0.445 |
| 62.50 | 0.414 | 0.320 |
| 31.25 | 0.241 | 0.147 |
| 0.00 | 0.094 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 95-107% | 101% |
| EDTA plasma(n=5) | 80-93% | 86% |
| Heparin plasma(n=5) | 94-105% | 98% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 86-95% | 93-103% | 87-98% | 85-96% |
| EDTA plasma(n=5) | 88-97% | 86-102% | 91-103% | 89-97% |
| Heparin plasma(n=5) | 85-102% | 87-96% | 82-90% | 92-101% |