| 48 T | $320.00 |
| 96 T | $458.00 |
| 96 T*5 | $1947.00 |
| Concentration (µg/mL) | OD | Corrected OD |
|---|---|---|
| 20.00 | 2.154 | 2.059 |
| 10.00 | 1.896 | 1.801 |
| 5.00 | 1.276 | 1.181 |
| 2.50 | 0.958 | 0.863 |
| 1.25 | 0.617 | 0.522 |
| 0.63 | 0.172 | 0.077 |
| 0.32 | 0.226 | 0.131 |
| 0.00 | 0.095 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 87-99% | 93% |
| EDTA plasma(n=5) | 92-105% | 98% |
| Heparin plasma(n=5) | 83-95% | 89% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 85-92% | 79-96% | 89-100% | 85-96% |
| EDTA plasma(n=5) | 85-94% | 87-103% | 82-94% | 78-90% |
| Heparin plasma(n=5) | 78-102% | 83-96% | 79-93% | 88-97% |