Size: | Price: |
---|---|
48 T | $265.00 |
96 T | $376.00 |
Concentration (ng/mL) | OD | Corrected OD |
---|---|---|
10.00 | 2.266 | 2.178 |
5.00 | 1.612 | 1.524 |
2.50 | 1.106 | 1.018 |
1.25 | 0.892 | 0.804 |
0.63 | 0.567 | 0.479 |
0.32 | 0.361 | 0.273 |
0.16 | 0.258 | 0.170 |
0.00 | 0.088 | 0.000 |
Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
Matrix | Recovery range | Average |
---|---|---|
serum(n=5) | 85-97% | 91% |
EDTA plasma(n=5) | 82-94% | 88% |
Heparin plasma(n=5) | 90-103% | 96% |
Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
---|---|---|---|---|
serum(n=5) | 95-106% | 85-104% | 90-98% | 85-101% |
EDTA plasma(n=5) | 81-99% | 85-92% | 82-96% | 82-101% |
Heparin plasma(n=5) | 87-96% | 93-108% | 80-101% | 87-96% |