Products

Horse HSP-70/HSPA9(HeatShock Protein 70) ELISA Kit
ELK9389
Size: Price:
48 T $412.00
96 T $588.00

Overview

Product name: Horse HSP-70/HSPA9(HeatShock Protein 70) ELISA Kit
Reactivity: Horse
Alternative Names: DnaK; 70 Kilodalton Heat Shock Proteins;Heat shock 70 kDa protein 1; 2; heat shock 70kDa protein 1A; HSP70; HSP70 1; HSP70 1; HSP70 2; HSP70 1A; HSP70.1; HSP70.2; HSP70I; HSP72; HSPA1; HSPA1A; HSPA1B; HSP70; Heat Shock Protein 70
Assay Type: Sandwich
Sensitivity: 0.55 ng/mL
Standard: 100 ng/mL
Detection Range: 1.57-100 ng/mL
Sample Type: Serum, plasma, tissue homogenates, cell lysates, cell culture supernates and other biological fluids
Assay Length: 3.5h
Research Area: Tumor immunity;Infection immunity;Developmental science;
Test principle: The test principle applied in this kit is Sandwich enzyme immunoassay. The microtiter plate provided in this kit has been pre-coated with an antibody specific to Horse HSP-70/HSPA9. Standards or samples are added to the appropriate microtiter plate wells then with a biotin-conjugated antibody specific to Horse HSP-70/HSPA9. Next, Avidin conjugated to Horseradish Peroxidase (HRP) is added to each microplate well and incubated. After TMB substrate solution is added, only those wells that contain Horse HSP-70/HSPA9, biotin-conjugated antibody and enzyme-conjugated Avidin will exhibit a change in color. The enzyme-substrate reaction is terminated by the addition of sulphuric acid solution and the color change is measured spectrophotometrically at a wavelength of 450nm ± 10nm. The concentration of Horse HSP-70/HSPA9 in the samples is then determined by comparing the OD of the samples to the standard curve.

Standard curve

Concentration (ng/mL) OD Corrected OD
100.00 2.268 2.172
50.00 1.607 1.511
25.00 1.196 1.100
12.50 0.828 0.732
6.25 0.528 0.432
3.13 0.389 0.293
1.57 0.194 0.098
0.00 0.096 0.000

Precision

Intra-assay Precision (Precision within an assay):CV%<8%

Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.

Inter-assay Precision (Precision between assays):CV%<10%

Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.

Recovery

Matrices listed below were spiked with certain level of recombinant HSP-70/HSPA9 and the recovery rates were calculated by comparing the measured value to the expected amount of HSP-70/HSPA9 in samples.
Matrix Recovery range Average
serum(n=5) 88-102% 95%
EDTA plasma(n=5) 82-94% 88%
Heparin plasma(n=5) 78-91% 84%

Linearity

The linearity of the kit was assayed by testing samples spiked with appropriate concentration of HSP-70/HSPA9 and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.
Matrix 1:2 1:4 1:8 1:16
serum(n=5) 92-101% 82-98% 78-102% 83-96%
EDTA plasma(n=5) 79-93% 88-97% 95-104% 93-102%
Heparin plasma(n=5) 89-96% 90-99% 95-106% 87-101%