Size: | Price: |
---|---|
48 T | $320.00 |
96 T | $458.00 |
Concentration (ng/mL) | OD | Corrected OD |
---|---|---|
50.00 | 2.401 | 2.315 |
25.00 | 1.675 | 1.589 |
12.50 | 1.294 | 1.208 |
6.25 | 0.933 | 0.847 |
3.13 | 0.611 | 0.525 |
1.57 | 0.352 | 0.266 |
0.79 | 0.238 | 0.152 |
0.00 | 0.086 | 0.000 |
Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
Matrix | Recovery range | Average |
---|---|---|
serum(n=5) | 95-107% | 101% |
EDTA plasma(n=5) | 86-99% | 92% |
Heparin plasma(n=5) | 89-103% | 96% |
Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
---|---|---|---|---|
serum(n=5) | 82-94% | 85-96% | 78-90% | 91-99% |
EDTA plasma(n=5) | 80-92% | 86-102% | 87-103% | 85-94% |
Heparin plasma(n=5) | 79-96% | 92-105% | 91-99% | 85-102% |