Size: | Price: |
---|---|
48 T | $320.00 |
96 T | $458.00 |
Concentration (pg/mL) | OD | Corrected OD |
---|---|---|
4000.00 | 2.054 | 1.969 |
2000.00 | 1.527 | 1.442 |
1000.00 | 1.142 | 1.057 |
500.00 | 0.831 | 0.746 |
250.00 | 0.496 | 0.411 |
125.00 | 0.357 | 0.272 |
62.50 | 0.189 | 0.104 |
0.00 | 0.085 | 0.000 |
Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
Matrix | Recovery range | Average |
---|---|---|
serum(n=5) | 90-105% | 97% |
EDTA plasma(n=5) | 87-99% | 93% |
Heparin plasma(n=5) | 82-95% | 88% |
Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
---|---|---|---|---|
serum(n=5) | 87-95% | 78-91% | 97-103% | 95-101% |
EDTA plasma(n=5) | 86-93% | 91-101% | 87-98% | 95-104% |
Heparin plasma(n=5) | 88-101% | 93-106% | 85-95% | 92-104% |