Size: | Price: |
---|---|
48 T | $412.00 |
96 T | $588.00 |
Concentration (ng/mL) | OD | Corrected OD |
---|---|---|
10.00 | 1.952 | 1.870 |
5.00 | 1.602 | 1.520 |
2.50 | 1.087 | 1.005 |
1.25 | 0.839 | 0.757 |
0.63 | 0.529 | 0.447 |
0.32 | 0.323 | 0.241 |
0.16 | 0.189 | 0.107 |
0.00 | 0.082 | 0.000 |
Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
Matrix | Recovery range | Average |
---|---|---|
serum(n=5) | 82-95% | 88% |
EDTA plasma(n=5) | 94-107% | 100% |
Heparin plasma(n=5) | 87-99% | 93% |
Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
---|---|---|---|---|
serum(n=5) | 87-102% | 82-91% | 86-93% | 81-94% |
EDTA plasma(n=5) | 89-98% | 90-99% | 95-106% | 87-101% |
Heparin plasma(n=5) | 86-93% | 81-94% | 83-96% | 97-103% |