Products

Plant GA(Gibberellic Acid) ELISA Kit
ELK8754
Size: Price:
48 T $265.00
96 T $376.00

Overview

Product name: Plant GA(Gibberellic Acid) ELISA Kit
Reactivity: Plant
Alternative Names: GA3; Gibberellin A3; Gibberellins
Assay Type: Competitive Inhibition
Sensitivity: 52.1 ng/mL
Standard: 10000 ng/mL
Detection Range: 156.25-10000 ng/mL
Sample Type: plant tissues
Assay Length: 2h
Research Area: Nutrition metabolism;Hormone metabolism;
Test principle: This assay employs the competitive inhibition enzyme immunoassay technique. The microtiter plate provided in this kit has been pre-coated with Plant GA protein. Standards or samples are added to the appropriate microtiter plate wells then with a biotin-conjugated antibody specific to Plant GA. Next, Avidin conjugated to Horseradish Peroxidase (HRP) is added to each microplate well and incubated. After TMB substrate solution is added. The enzyme-substrate reaction is terminated by the addition of sulphuric acid solution and the color change is measured spectrophotometrically at a wavelength of 450nm ± 10nm. The concentration of Plant GA in the samples is then determined by comparing the OD of the samples to the standard curve.

Standard curve

Concentration (ng/mL) OD Corrected OD
10000.00 0.191
5000.00 0.386
2500.00 0.689
1250.00 0.895
625.00 1.275
312.50 1.596
156.25 1.724
0.00 2.375

Precision

Intra-assay Precision (Precision within an assay):CV%<8%

Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.

Inter-assay Precision (Precision between assays):CV%<10%

Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.

Recovery

Matrices listed below were spiked with certain level of recombinant GA and the recovery rates were calculated by comparing the measured value to the expected amount of GA in samples.
Matrix Recovery range Average
serum(n=5) 80-95% 87%
EDTA plasma(n=5) 86-99% 92%
Heparin plasma(n=5) 82-95% 97%

Linearity

The linearity of the kit was assayed by testing samples spiked with appropriate concentration of GA and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.
Matrix 1:2 1:4 1:8 1:16
serum(n=5) 96-101% 85-92% 79-96% 89-100%
EDTA plasma(n=5) 91-102% 89-97% 86-102% 94-106%
Heparin plasma(n=5) 93-102% 91-107% 87-101% 95-102%