Size: | Price: |
---|---|
48 T | $320.00 |
96 T | $458.00 |
Concentration (ng/mL) | OD | Corrected OD |
---|---|---|
10.00 | 1.964 | 1.885 |
5.00 | 1.602 | 1.523 |
2.50 | 1.073 | 0.994 |
1.25 | 0.866 | 0.787 |
0.63 | 0.527 | 0.448 |
0.32 | 0.324 | 0.245 |
0.16 | 0.194 | 0.115 |
0.00 | 0.079 | 0.000 |
Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
Matrix | Recovery range | Average |
---|---|---|
serum(n=5) | 80-95% | 87% |
EDTA plasma(n=5) | 84-102% | 93% |
Heparin plasma(n=5) | 87-95% | 90% |
Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
---|---|---|---|---|
serum(n=5) | 86-98% | 91-103% | 85-97% | 87-96% |
EDTA plasma(n=5) | 88-105% | 79-93% | 87-98% | 85-96% |
Heparin plasma(n=5) | 95-103% | 80-96% | 87-98% | 85-92% |