Size: | Price: |
---|---|
48 T | $320.00 |
96 T | $458.00 |
Concentration (pg/mL) | OD | Corrected OD |
---|---|---|
2000.00 | 1.951 | 1.855 |
1000.00 | 1.682 | 1.586 |
500.00 | 1.151 | 1.055 |
250.00 | 0.861 | 0.765 |
125.00 | 0.559 | 0.463 |
62.50 | 0.319 | 0.223 |
31.25 | 0.225 | 0.129 |
0.00 | 0.096 | 0.000 |
Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
Matrix | Recovery range | Average |
---|---|---|
serum(n=5) | 78-95% | 86% |
EDTA plasma(n=5) | 80-92% | 86% |
Heparin plasma(n=5) | 80-96% | 88% |
Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
---|---|---|---|---|
serum(n=5) | 90-99% | 83-97% | 81-103% | 89-101% |
EDTA plasma(n=5) | 87-98% | 92-105% | 95-107% | 82-96% |
Heparin plasma(n=5) | 79-91% | 89-103% | 82-95% | 86-99% |