Products

Human Bisphenol A ELISA Kit
ELK8383
Size: Price:
48 T $320.00
96 T $458.00

OverviewPublications

Product name: Human Bisphenol A ELISA Kit
Reactivity: Human
Alternative Names: BPA
Assay Type: Sandwich
Sensitivity: 0.52 ng/mL
Standard: 80 ng/mL
Detection Range: 1.25-80 ng/mL
Sample Type: Plasma, Serum,
Assay Length: 3.5h
Research Area: Neuro science;
Test principle: The test principle applied in this kit is Sandwich enzyme immunoassay. The microtiter plate provided in this kit has been pre-coated with an antibody specific to Human Bisphenol A. Standards or samples are added to the appropriate microtiter plate wells then with a biotin-conjugated antibody specific to Human Bisphenol A. Next, Avidin conjugated to Horseradish Peroxidase (HRP) is added to each microplate well and incubated. After TMB substrate solution is added, only those wells that contain Human Bisphenol A, biotin-conjugated antibody and enzyme-conjugated Avidin will exhibit a change in color. The enzyme-substrate reaction is terminated by the addition of sulphuric acid solution and the color change is measured spectrophotometrically at a wavelength of 450nm ± 10nm. The concentration of Human Bisphenol A in the samples is then determined by comparing the OD of the samples to the standard curve.

Standard curve

Concentration (ng/mL) OD Corrected OD
80.00 2.258 2.165
40.00 1.562 1.469
20.00 1.073 0.980
10.00 0.892 0.799
5.00 0.593 0.500
2.50 0.318 0.225
1.25 0.216 0.123
0.00 0.093 0.000

Precision

Intra-assay Precision (Precision within an assay):CV%<8%

Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.

Inter-assay Precision (Precision between assays):CV%<10%

Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.

Recovery

Matrices listed below were spiked with certain level of recombinant Bisphenol A and the recovery rates were calculated by comparing the measured value to the expected amount of Bisphenol A in samples.
Matrix Recovery range Average
serum(n=5) 90-107% 98%
EDTA plasma(n=5) 87-99% 93%
Heparin plasma(n=5) 82-98% 90%

Linearity

The linearity of the kit was assayed by testing samples spiked with appropriate concentration of Bisphenol A and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.
Matrix 1:2 1:4 1:8 1:16
serum(n=5) 85-96% 86-98% 92-103% 87-93%
EDTA plasma(n=5) 85-93% 96-103% 97-105% 89-97%
Heparin plasma(n=5) 84-102% 87-103% 97-104% 82-97%