Size: | Price: |
---|---|
48 T | $412.00 |
96 T | $588.00 |
Concentration (ng/mL) | OD | Corrected OD |
---|---|---|
10.00 | 2.291 | 2.218 |
5.00 | 1.642 | 1.569 |
2.50 | 1.196 | 1.123 |
1.25 | 0.963 | 0.890 |
0.63 | 0.548 | 0.475 |
0.32 | 0.322 | 0.249 |
0.16 | 0.179 | 0.106 |
0.00 | 0.073 | 0.000 |
Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
Matrix | Recovery range | Average |
---|---|---|
serum(n=5) | 87-99% | 93% |
EDTA plasma(n=5) | 87-95% | 91% |
Heparin plasma(n=5) | 93-107% | 100% |
Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
---|---|---|---|---|
serum(n=5) | 86-93% | 91-105% | 87-99% | 99-105% |
EDTA plasma(n=5) | 95-102% | 97-105% | 92-101% | 85-92% |
Heparin plasma(n=5) | 88-97% | 79-92% | 85-93% | 89-97% |