Products

Cattle PEDF(Pigment Epithelium Derived Factor) ELISA Kit
ELK7999
Size: Price:
48 T $412.00
96 T $588.00

Overview

Product name: Cattle PEDF(Pigment Epithelium Derived Factor) ELISA Kit
Reactivity: Cattle
Alternative Names: SERPINF1; EPC1; PIG35; SDF3; Serpin F1; Cell proliferation-inducing gene 35; Serpin Peptidase Inhibitor,Clade F Member 1(Alpha-2 Antiplasmin); Stromal Cell Derived Factor 3
Assay Type: Sandwich
Sensitivity: 0.135 ng/mL
Standard: 20 ng/mL
Detection Range: 0.32-20 ng/mL
Sample Type: Serum, plasma and other biological fluids.
Assay Length: 3.5h
Research Area: Cytokine;Neuro science;
Uniprot ID: Q95121
Test principle: The test principle applied in this kit is Sandwich enzyme immunoassay. The microtiter plate provided in this kit has been pre-coated with an antibody specific to Cattle PEDF. Standards or samples are added to the appropriate microtiter plate wells then with a biotin-conjugated antibody specific to Cattle PEDF. Next, Avidin conjugated to Horseradish Peroxidase (HRP) is added to each microplate well and incubated. After TMB substrate solution is added, only those wells that contain Cattle PEDF, biotin-conjugated antibody and enzyme-conjugated Avidin will exhibit a change in color. The enzyme-substrate reaction is terminated by the addition of sulphuric acid solution and the color change is measured spectrophotometrically at a wavelength of 450nm ± 10nm. The concentration of Cattle PEDF in the samples is then determined by comparing the OD of the samples to the standard curve.

Standard curve

Concentration (ng/mL) OD Corrected OD
20.00 2.255 2.166
10.00 1.597 1.508
5.00 1.182 1.093
2.50 0.752 0.663
1.25 0.572 0.483
0.63 0.355 0.266
0.32 0.182 0.093
0.00 0.089 0.000

Precision

Intra-assay Precision (Precision within an assay):CV%<8%

Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.

Inter-assay Precision (Precision between assays):CV%<10%

Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.

Recovery

Matrices listed below were spiked with certain level of recombinant PEDF and the recovery rates were calculated by comparing the measured value to the expected amount of PEDF in samples.
Matrix Recovery range Average
serum(n=5) 85-97% 91%
EDTA plasma(n=5) 85-97% 91%
Heparin plasma(n=5) 83-99% 95%

Linearity

The linearity of the kit was assayed by testing samples spiked with appropriate concentration of PEDF and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.
Matrix 1:2 1:4 1:8 1:16
serum(n=5) 87-92% 85-94% 89-98% 85-94%
EDTA plasma(n=5) 89-102% 82-94% 85-96% 78-90%
Heparin plasma(n=5) 93-102% 80-93% 91-100% 88-95%