Size: | Price: |
---|---|
48 T | $320.00 |
96 T | $458.00 |
Concentration (ng/mL) | OD | Corrected OD |
---|---|---|
10.00 | 2.126 | 2.025 |
5.00 | 1.613 | 1.512 |
2.50 | 1.073 | 0.972 |
1.25 | 0.757 | 0.656 |
0.63 | 0.468 | 0.367 |
0.32 | 0.361 | 0.260 |
0.16 | 0.192 | 0.091 |
0.00 | 0.101 | 0.000 |
Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
Matrix | Recovery range | Average |
---|---|---|
serum(n=5) | 78-96% | 87% |
EDTA plasma(n=5) | 86-99% | 92% |
Heparin plasma(n=5) | 89-103% | 96% |
Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
---|---|---|---|---|
serum(n=5) | 95-106% | 83-96% | 87-98% | 78-89% |
EDTA plasma(n=5) | 86-102% | 87-103% | 85-101% | 81-99% |
Heparin plasma(n=5) | 95-104% | 93-105% | 82-101% | 91-103% |