Size: | Price: |
---|---|
48 T | $412.00 |
96 T | $588.00 |
Concentration (ng/mL) | OD | Corrected OD |
---|---|---|
100.00 | 2.052 | 1.969 |
50.00 | 1.507 | 1.424 |
25.00 | 1.240 | 1.157 |
12.50 | 0.839 | 0.756 |
6.25 | 0.573 | 0.490 |
3.13 | 0.363 | 0.280 |
1.57 | 0.193 | 0.110 |
0.00 | 0.083 | 0.000 |
Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
Matrix | Recovery range | Average |
---|---|---|
serum(n=5) | 78-95% | 86% |
EDTA plasma(n=5) | 80-96% | 88% |
Heparin plasma(n=5) | 80-92% | 86% |
Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
---|---|---|---|---|
serum(n=5) | 89-99% | 88-104% | 84-98% | 95-105% |
EDTA plasma(n=5) | 95-104% | 87-96% | 92-101% | 82-90% |
Heparin plasma(n=5) | 78-92% | 83-96% | 87-98% | 95-102% |