Size: | Price: |
---|---|
48 T | $320.00 |
96 T | $458.00 |
Concentration (pg/mL) | OD | Corrected OD |
---|---|---|
200.00 | 0.182 | |
100.00 | 0.411 | |
50.00 | 0.692 | |
25.00 | 0.933 | |
12.50 | 1.253 | |
6.25 | 1.587 | |
3.13 | 1.813 | |
0.00 | 2.211 |
Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
Matrix | Recovery range | Average |
---|---|---|
serum(n=5) | 82-95% | 88% |
EDTA plasma(n=5) | 89-103% | 96% |
Heparin plasma(n=5) | 78-93% | 90% |
Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
---|---|---|---|---|
serum(n=5) | 87-102% | 82-91% | 85-94% | 87-101% |
EDTA plasma(n=5) | 86-97% | 88-95% | 92-101% | 87-98% |
Heparin plasma(n=5) | 80-101% | 87-101% | 86-97% | 88-95% |