Products

Rat PDGFRb(Platelet Derived Growth Factor Receptor Beta) ELISA Kit
ELK7580
Size: Price:
48 T $320.00
96 T $458.00

Overview

Product name: Rat PDGFRb(Platelet Derived Growth Factor Receptor Beta) ELISA Kit
Reactivity: Rat
Alternative Names: CD140b; JTK12; PDGF-R-Beta; PDGFR; PDGFR1; CD140 antigen-like family member B; Platelet-derived growth factor receptor 1; Beta-type platelet-derived growth factor receptor
Assay Type: Sandwich
Sensitivity: 0.062 ng/mL
Standard: 10 ng/mL
Detection Range: 0.16-10 ng/mL
Sample Type: Tissue homogenates, cell lysates and other biological fluids
Assay Length: 3.5h
Research Area: Signal transduction;Tumor immunity;Infection immunity;Hematology;
Uniprot ID: Q05030
Test principle: The test principle applied in this kit is Sandwich enzyme immunoassay. The microtiter plate provided in this kit has been pre-coated with an antibody specific to Rat PDGFRb. Standards or samples are added to the appropriate microtiter plate wells then with a biotin-conjugated antibody specific to Rat PDGFRb. Next, Avidin conjugated to Horseradish Peroxidase (HRP) is added to each microplate well and incubated. After TMB substrate solution is added, only those wells that contain Rat PDGFRb, biotin-conjugated antibody and enzyme-conjugated Avidin will exhibit a change in color. The enzyme-substrate reaction is terminated by the addition of sulphuric acid solution and the color change is measured spectrophotometrically at a wavelength of 450nm ± 10nm. The concentration of Rat PDGFRb in the samples is then determined by comparing the OD of the samples to the standard curve.

Standard curve

Concentration (ng/mL) OD Corrected OD
10.00 2.059 1.962
5.00 1.328 1.231
2.50 1.034 0.937
1.25 0.828 0.731
0.63 0.559 0.462
0.32 0.395 0.298
0.16 0.241 0.144
0.00 0.097 0.000

Precision

Intra-assay Precision (Precision within an assay):CV%<8%

Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.

Inter-assay Precision (Precision between assays):CV%<10%

Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.

Recovery

Matrices listed below were spiked with certain level of recombinant PDGFRb and the recovery rates were calculated by comparing the measured value to the expected amount of PDGFRb in samples.
Matrix Recovery range Average
serum(n=5) 87-99% 93%
EDTA plasma(n=5) 93-107% 100%
Heparin plasma(n=5) 85-99% 92%

Linearity

The linearity of the kit was assayed by testing samples spiked with appropriate concentration of PDGFRb and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.
Matrix 1:2 1:4 1:8 1:16
serum(n=5) 85-96% 86-95% 90-102% 81-90%
EDTA plasma(n=5) 79-91% 95-104% 93-102% 79-94%
Heparin plasma(n=5) 87-98% 92-104% 82-96% 95-102%