Size: | Price: |
---|---|
48 T | $320.00 |
96 T | $458.00 |
Concentration (ng/mL) | OD | Corrected OD |
---|---|---|
100.00 | 2.242 | 2.147 |
50.00 | 1.779 | 1.684 |
25.00 | 1.106 | 1.011 |
12.50 | 0.828 | 0.733 |
6.25 | 0.582 | 0.487 |
3.13 | 0.315 | 0.220 |
1.57 | 0.174 | 0.079 |
0.00 | 0.095 | 0.000 |
Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
Matrix | Recovery range | Average |
---|---|---|
serum(n=5) | 80-93% | 86% |
EDTA plasma(n=5) | 95-107% | 101% |
Heparin plasma(n=5) | 80-95% | 87% |
Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
---|---|---|---|---|
serum(n=5) | 93-102% | 80-93% | 91-100% | 88-95% |
EDTA plasma(n=5) | 92-107% | 83-99% | 80-96% | 91-101% |
Heparin plasma(n=5) | 85-97% | 86-102% | 87-99% | 99-105% |