Size: | Price: |
---|---|
48 T | $320.00 |
96 T | $458.00 |
Concentration (ng/mL) | OD | Corrected OD |
---|---|---|
60.00 | 2.168 | 2.066 |
30.00 | 1.629 | 1.527 |
15.00 | 1.175 | 1.073 |
7.50 | 0.895 | 0.793 |
3.75 | 0.553 | 0.451 |
1.88 | 0.366 | 0.264 |
0.94 | 0.263 | 0.161 |
0.00 | 0.102 | 0.000 |
Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
Matrix | Recovery range | Average |
---|---|---|
serum(n=5) | 80-92% | 86% |
EDTA plasma(n=5) | 83-97% | 90% |
Heparin plasma(n=5) | 86-99% | 92% |
Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
---|---|---|---|---|
serum(n=5) | 88-101% | 93-106% | 85-95% | 92-104% |
EDTA plasma(n=5) | 86-97% | 88-95% | 92-101% | 87-98% |
Heparin plasma(n=5) | 85-92% | 79-96% | 89-100% | 85-96% |