Size: | Price: |
---|---|
48 T | $320.00 |
96 T | $458.00 |
Concentration (ng/mL) | OD | Corrected OD |
---|---|---|
10.00 | 1.948 | 1.859 |
5.00 | 1.586 | 1.497 |
2.50 | 1.132 | 1.043 |
1.25 | 0.812 | 0.723 |
0.63 | 0.506 | 0.417 |
0.32 | 0.375 | 0.286 |
0.16 | 0.206 | 0.117 |
0.00 | 0.089 | 0.000 |
Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
Matrix | Recovery range | Average |
---|---|---|
serum(n=5) | 96-107% | 101% |
EDTA plasma(n=5) | 87-99% | 92% |
Heparin plasma(n=5) | 84-96% | 90% |
Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
---|---|---|---|---|
serum(n=5) | 78-92% | 83-96% | 95-107% | 87-102% |
EDTA plasma(n=5) | 78-96% | 89-102% | 85-97% | 79-97% |
Heparin plasma(n=5) | 92-104% | 87-96% | 89-101% | 82-90% |