Size: | Price: |
---|---|
48 T | $320.00 |
96 T | $458.00 |
Concentration (ng/mL) | OD | Corrected OD |
---|---|---|
50.00 | 1.951 | 1.856 |
25.00 | 1.522 | 1.427 |
12.50 | 1.088 | 0.993 |
6.25 | 0.805 | 0.710 |
3.13 | 0.506 | 0.411 |
1.57 | 0.317 | 0.222 |
0.79 | 0.176 | 0.081 |
0.00 | 0.095 | 0.000 |
Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
Matrix | Recovery range | Average |
---|---|---|
serum(n=5) | 82-94% | 88% |
EDTA plasma(n=5) | 97-107% | 102% |
Heparin plasma(n=5) | 87-99% | 93% |
Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
---|---|---|---|---|
serum(n=5) | 79-96% | 85-99% | 79-93% | 89-97% |
EDTA plasma(n=5) | 88-103% | 86-92% | 91-99% | 95-102% |
Heparin plasma(n=5) | 79-91% | 87-96% | 83-101% | 82-95% |