Size: | Price: |
---|---|
48 T | $320.00 |
96 T | $458.00 |
Concentration (ng/mL) | OD | Corrected OD |
---|---|---|
50.00 | 2.166 | 2.072 |
25.00 | 1.790 | 1.696 |
12.50 | 1.239 | 1.145 |
6.25 | 0.782 | 0.688 |
3.13 | 0.562 | 0.468 |
1.57 | 0.361 | 0.267 |
0.79 | 0.192 | 0.098 |
0.00 | 0.094 | 0.000 |
Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
Matrix | Recovery range | Average |
---|---|---|
serum(n=5) | 87-99% | 93% |
EDTA plasma(n=5) | 80-92% | 86% |
Heparin plasma(n=5) | 92-106% | 99% |
Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
---|---|---|---|---|
serum(n=5) | 95-103% | 86-92% | 89-97% | 91-103% |
EDTA plasma(n=5) | 78-92% | 79-93% | 94-106% | 83-97% |
Heparin plasma(n=5) | 83-95% | 88-102% | 85-97% | 80-92% |