Size: | Price: |
---|---|
48 T | $320.00 |
96 T | $458.00 |
Concentration (ng/mL) | OD | Corrected OD |
---|---|---|
10.00 | 2.096 | 2.010 |
5.00 | 1.577 | 1.491 |
2.50 | 1.121 | 1.035 |
1.25 | 0.801 | 0.715 |
0.63 | 0.474 | 0.388 |
0.32 | 0.372 | 0.286 |
0.16 | 0.129 | 0.043 |
0.00 | 0.086 | 0.000 |
Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
Matrix | Recovery range | Average |
---|---|---|
serum(n=5) | 80-95% | 87% |
EDTA plasma(n=5) | 80-95% | 87% |
Heparin plasma(n=5) | 86-99% | 92% |
Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
---|---|---|---|---|
serum(n=5) | 83-97% | 82-96% | 86-98% | 93-101% |
EDTA plasma(n=5) | 87-103% | 79-92% | 86-99% | 84-97% |
Heparin plasma(n=5) | 88-97% | 95-106% | 83-98% | 93-105% |