48 T | $320.00 |
96 T | $458.00 |
Concentration (ng/mL) | OD | Corrected OD |
---|---|---|
10.00 | 2.119 | 2.034 |
5.00 | 1.527 | 1.442 |
2.50 | 1.069 | 0.984 |
1.25 | 0.847 | 0.762 |
0.63 | 0.527 | 0.442 |
0.32 | 0.324 | 0.239 |
0.16 | 0.258 | 0.173 |
0.00 | 0.085 | 0.000 |
Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
Matrix | Recovery range | Average |
---|---|---|
serum(n=5) | 92-105% | 98% |
EDTA plasma(n=5) | 85-97% | 91% |
Heparin plasma(n=5) | 78-92% | 85% |
Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
---|---|---|---|---|
serum(n=5) | 85-98% | 81-96% | 87-98% | 92-103% |
EDTA plasma(n=5) | 89-99% | 88-104% | 84-98% | 95-105% |
Heparin plasma(n=5) | 86-97% | 83-96% | 91-103% | 96-105% |