Size: | Price: |
---|---|
48 T | $320.00 |
96 T | $458.00 |
Concentration (ng/mL) | OD | Corrected OD |
---|---|---|
20.00 | 2.153 | 2.064 |
10.00 | 1.577 | 1.488 |
5.00 | 1.121 | 1.032 |
2.50 | 0.728 | 0.639 |
1.25 | 0.547 | 0.458 |
0.63 | 0.357 | 0.268 |
0.32 | 0.179 | 0.090 |
0.00 | 0.089 | 0.000 |
Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
Matrix | Recovery range | Average |
---|---|---|
serum(n=5) | 97-105% | 101% |
EDTA plasma(n=5) | 95-107% | 101% |
Heparin plasma(n=5) | 90-105% | 97% |
Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
---|---|---|---|---|
serum(n=5) | 88-95% | 98-107% | 90-99% | 88-97% |
EDTA plasma(n=5) | 85-97% | 84-100% | 95-103% | 88-104% |
Heparin plasma(n=5) | 86-95% | 91-106% | 87-100% | 95-101% |