Size: | Price: |
---|---|
48 T | $320.00 |
96 T | $458.00 |
Concentration (pg/mL) | OD | Corrected OD |
---|---|---|
2000.00 | 2.126 | 2.022 |
1000.00 | 1.782 | 1.678 |
500.00 | 1.297 | 1.193 |
250.00 | 0.872 | 0.768 |
125.00 | 0.529 | 0.425 |
62.50 | 0.327 | 0.223 |
31.25 | 0.265 | 0.161 |
0.00 | 0.104 | 0.000 |
Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
Matrix | Recovery range | Average |
---|---|---|
serum(n=5) | 82-95% | 88% |
EDTA plasma(n=5) | 80-95% | 87% |
Heparin plasma(n=5) | 86-99% | 92% |
Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
---|---|---|---|---|
serum(n=5) | 83-96% | 86-95% | 90-102% | 81-93% |
EDTA plasma(n=5) | 87-96% | 92-105% | 89-102% | 92-101% |
Heparin plasma(n=5) | 85-96% | 87-101% | 88-102% | 87-102% |