Size: | Price: |
---|---|
48 T | $320.00 |
96 T | $458.00 |
Concentration (ng/mL) | OD | Corrected OD |
---|---|---|
10.00 | 1.998 | 1.902 |
5.00 | 1.700 | 1.604 |
2.50 | 1.043 | 0.947 |
1.25 | 0.756 | 0.660 |
0.63 | 0.518 | 0.422 |
0.32 | 0.351 | 0.255 |
0.16 | 0.228 | 0.132 |
0.00 | 0.096 | 0.000 |
Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
Matrix | Recovery range | Average |
---|---|---|
serum(n=5) | 89-103% | 96% |
EDTA plasma(n=5) | 85-97% | 91% |
Heparin plasma(n=5) | 80-95% | 87% |
Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
---|---|---|---|---|
serum(n=5) | 88-96% | 86-95% | 90-103% | 87-90% |
EDTA plasma(n=5) | 85-102% | 87-96% | 92-101% | 89-104% |
Heparin plasma(n=5) | 95-102% | 87-96% | 78-92% | 85-97% |