Size: | Price: |
---|---|
48 T | $412.00 |
96 T | $588.00 |
Concentration (ng/mL) | OD | Corrected OD |
---|---|---|
200.00 | 2.126 | 2.025 |
100.00 | 1.593 | 1.492 |
50.00 | 1.282 | 1.181 |
25.00 | 0.796 | 0.695 |
12.50 | 0.539 | 0.438 |
6.25 | 0.365 | 0.264 |
3.13 | 0.202 | 0.101 |
0.00 | 0.101 | 0.000 |
Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
Matrix | Recovery range | Average |
---|---|---|
serum(n=5) | 80-96% | 88% |
EDTA plasma(n=5) | 80-92% | 86% |
Heparin plasma(n=5) | 78-95% | 86% |
Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
---|---|---|---|---|
serum(n=5) | 92-105% | 82-92% | 86-94% | 81-95% |
EDTA plasma(n=5) | 92-105% | 83-92% | 89-98% | 85-94% |
Heparin plasma(n=5) | 89-99% | 83-95% | 98-104% | 96-106% |