Size: | Price: |
---|---|
48 T | $320.00 |
96 T | $458.00 |
Concentration (pg/mL) | OD | Corrected OD |
---|---|---|
1000.00 | 1.966 | 1.863 |
500.00 | 1.585 | 1.482 |
250.00 | 1.115 | 1.012 |
125.00 | 0.882 | 0.779 |
62.50 | 0.535 | 0.432 |
31.25 | 0.368 | 0.265 |
15.63 | 0.236 | 0.133 |
0.00 | 0.103 | 0.000 |
Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
Matrix | Recovery range | Average |
---|---|---|
serum(n=5) | 84-97% | 90% |
EDTA plasma(n=5) | 87-95% | 91% |
Heparin plasma(n=5) | 78-90% | 84% |
Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
---|---|---|---|---|
serum(n=5) | 87-96% | 92-105% | 89-102% | 92-101% |
EDTA plasma(n=5) | 91-98% | 79-93% | 92-101% | 82-90% |
Heparin plasma(n=5) | 85-92% | 95-103% | 97-105% | 93-102% |