Size: | Price: |
---|---|
48 T | $412.00 |
96 T | $588.00 |
Concentration (ng/mL) | OD | Corrected OD |
---|---|---|
25.00 | 2.045 | 1.948 |
12.50 | 1.528 | 1.431 |
6.25 | 1.185 | 1.088 |
3.13 | 0.875 | 0.778 |
1.57 | 0.557 | 0.460 |
0.79 | 0.325 | 0.228 |
0.40 | 0.213 | 0.116 |
0.00 | 0.097 | 0.000 |
Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
Matrix | Recovery range | Average |
---|---|---|
serum(n=5) | 84-106% | 100% |
EDTA plasma(n=5) | 87-93% | 90% |
Heparin plasma(n=5) | 80-90% | 85% |
Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
---|---|---|---|---|
serum(n=5) | 93-104% | 87-96% | 82-95% | 78-99% |
EDTA plasma(n=5) | 95-104% | 93-105% | 82-101% | 91-103% |
Heparin plasma(n=5) | 89-97% | 85-101% | 84-98% | 85-93% |