Products

Cattle TSP-1(Thrombospondin-1) ELISA Kit
ELK5945
Size: Price:
48 T $412.00
96 T $588.00

Overview

Product name: Cattle TSP-1(Thrombospondin-1) ELISA Kit
Reactivity: Cattle
Alternative Names: TSP1; Thrombospondin-1p180; THBS1; Thrombospondin 1
Assay Type: Sandwich
Sensitivity: 30 pg/mL
Standard: 5000 pg/mL
Detection Range: 78.13-5000 pg/mL
Sample Type: Serum and plasma
Assay Length: 3.5h
Research Area: Infection immunity;Hematology;
Uniprot ID: Q28178
Test principle: The test principle applied in this kit is Sandwich enzyme immunoassay. The microtiter plate provided in this kit has been pre-coated with an antibody specific to Cattle TSP-1. Standards or samples are added to the appropriate microtiter plate wells then with a biotin-conjugated antibody specific to Cattle TSP-1. Next, Avidin conjugated to Horseradish Peroxidase (HRP) is added to each microplate well and incubated. After TMB substrate solution is added, only those wells that contain Cattle TSP-1, biotin-conjugated antibody and enzyme-conjugated Avidin will exhibit a change in color. The enzyme-substrate reaction is terminated by the addition of sulphuric acid solution and the color change is measured spectrophotometrically at a wavelength of 450nm ± 10nm. The concentration of Cattle TSP-1 in the samples is then determined by comparing the OD of the samples to the standard curve.

Standard curve

Concentration (pg/mL) OD Corrected OD
5000.00 2.266 2.180
2500.00 1.643 1.557
1250.00 1.109 1.023
625.00 0.876 0.790
312.50 0.591 0.505
156.25 0.368 0.282
78.13 0.208 0.122
0.00 0.086 0.000

Precision

Intra-assay Precision (Precision within an assay):CV%<8%

Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.

Inter-assay Precision (Precision between assays):CV%<10%

Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.

Recovery

Matrices listed below were spiked with certain level of recombinant TSP-1 and the recovery rates were calculated by comparing the measured value to the expected amount of TSP-1 in samples.
Matrix Recovery range Average
serum(n=5) 87-99% 93%
EDTA plasma(n=5) 92-105% 98%
Heparin plasma(n=5) 83-95% 89%

Linearity

The linearity of the kit was assayed by testing samples spiked with appropriate concentration of TSP-1 and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.
Matrix 1:2 1:4 1:8 1:16
serum(n=5) 92-103% 83-98% 88-97% 96-104%
EDTA plasma(n=5) 79-93% 89-97% 85-94% 87-101%
Heparin plasma(n=5) 86-94% 93-102% 87-96% 92-105%