Size: | Price: |
---|---|
48 T | $412.00 |
96 T | $588.00 |
Concentration (pg/mL) | OD | Corrected OD |
---|---|---|
2000.00 | 1.908 | 1.809 |
1000.00 | 1.511 | 1.412 |
500.00 | 1.073 | 0.974 |
250.00 | 0.872 | 0.773 |
125.00 | 0.521 | 0.422 |
62.50 | 0.361 | 0.262 |
31.25 | 0.186 | 0.087 |
0.00 | 0.099 | 0.000 |
Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
Matrix | Recovery range | Average |
---|---|---|
serum(n=5) | 78-90% | 84% |
EDTA plasma(n=5) | 89-107% | 98% |
Heparin plasma(n=5) | 87-99% | 93% |
Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
---|---|---|---|---|
serum(n=5) | 85-92% | 83-96% | 93-102% | 87-98% |
EDTA plasma(n=5) | 89-102% | 92-101% | 88-102% | 87-102% |
Heparin plasma(n=5) | 93-104% | 83-96% | 95-103% | 88-95% |