Size: | Price: |
---|---|
48 T | $412.00 |
96 T | $588.00 |
Concentration (ng/mL) | OD | Corrected OD |
---|---|---|
10.00 | 2.266 | 2.170 |
5.00 | 1.602 | 1.506 |
2.50 | 1.185 | 1.089 |
1.25 | 0.875 | 0.779 |
0.63 | 0.518 | 0.422 |
0.32 | 0.303 | 0.207 |
0.16 | 0.186 | 0.090 |
0.00 | 0.096 | 0.000 |
Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
Matrix | Recovery range | Average |
---|---|---|
serum(n=5) | 83-98% | 91% |
EDTA plasma(n=5) | 91-99% | 95% |
Heparin plasma(n=5) | 82-95% | 88% |
Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
---|---|---|---|---|
serum(n=5) | 89-97% | 85-94% | 79-96% | 85-99% |
EDTA plasma(n=5) | 92-106% | 93-105% | 85-99% | 90-99% |
Heparin plasma(n=5) | 88-104% | 84-98% | 95-105% | 88-102% |