Products

Horse VEGFA(Vascular Endothelial Growth Factor A) ELISA Kit
ELK5807
Size: Price:
48 T $412.00
96 T $588.00

Overview

Product name: Horse VEGFA(Vascular Endothelial Growth Factor A) ELISA Kit
Reactivity: Horse
Alternative Names: VEGF; VEGF-A; VPF; Vascular Endothelial Cell Growth Factor A; Vascular Permeability Factor
Assay Type: Sandwich
Sensitivity: 5.5 pg/mL
Standard: 1000 pg/mL
Detection Range: 15.63-1000 pg/mL
Sample Type: Serum, plasma, tissue homogenates, cell lysates, cell culture supernates and other biological fluids.
Assay Length: 3.5h
Research Area: Cytokine;Apoptosis;Tumor immunity;
Uniprot ID: Q9GKR0
Test principle: The test principle applied in this kit is Sandwich enzyme immunoassay. The microtiter plate provided in this kit has been pre-coated with an antibody specific to Horse VEGFA. Standards or samples are added to the appropriate microtiter plate wells then with a biotin-conjugated antibody specific to Horse VEGFA. Next, Avidin conjugated to Horseradish Peroxidase (HRP) is added to each microplate well and incubated. After TMB substrate solution is added, only those wells that contain Horse VEGFA, biotin-conjugated antibody and enzyme-conjugated Avidin will exhibit a change in color. The enzyme-substrate reaction is terminated by the addition of sulphuric acid solution and the color change is measured spectrophotometrically at a wavelength of 450nm ± 10nm. The concentration of Horse VEGFA in the samples is then determined by comparing the OD of the samples to the standard curve.

Standard curve

Concentration (pg/mL) OD Corrected OD
1000.00 2.126 2.029
500.00 1.527 1.430
250.00 1.112 1.015
125.00 0.992 0.895
62.50 0.508 0.411
31.25 0.305 0.208
15.63 0.236 0.139
0.00 0.097 0.000

Precision

Intra-assay Precision (Precision within an assay):CV%<8%

Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.

Inter-assay Precision (Precision between assays):CV%<10%

Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.

Recovery

Matrices listed below were spiked with certain level of recombinant VEGFA and the recovery rates were calculated by comparing the measured value to the expected amount of VEGFA in samples.
Matrix Recovery range Average
serum(n=5) 83-97% 90%
EDTA plasma(n=5) 86-99% 92%
Heparin plasma(n=5) 89-103% 96%

Linearity

The linearity of the kit was assayed by testing samples spiked with appropriate concentration of VEGFA and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.
Matrix 1:2 1:4 1:8 1:16
serum(n=5) 88-97% 96-103% 86-92% 89-97%
EDTA plasma(n=5) 86-95% 91-106% 87-100% 95-101%
Heparin plasma(n=5) 95-102% 83-97% 88-96% 96-103%