Size: | Price: |
---|---|
48 T | $320.00 |
96 T | $458.00 |
Concentration (ng/mL) | OD | Corrected OD |
---|---|---|
100.00 | 1.882 | 1.775 |
50.00 | 1.582 | 1.475 |
25.00 | 1.123 | 1.016 |
12.50 | 0.875 | 0.768 |
6.25 | 0.632 | 0.525 |
3.13 | 0.354 | 0.247 |
1.57 | 0.228 | 0.121 |
0.00 | 0.107 | 0.000 |
Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
Matrix | Recovery range | Average |
---|---|---|
serum(n=5) | 80-93% | 86% |
EDTA plasma(n=5) | 93-107% | 100% |
Heparin plasma(n=5) | 87-99% | 93% |
Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
---|---|---|---|---|
serum(n=5) | 88-102% | 87-102% | 85-96% | 87-101% |
EDTA plasma(n=5) | 95-104% | 79-92% | 82-95% | 86-97% |