Size: | Price: |
---|---|
48 T | $320.00 |
96 T | $458.00 |
Concentration (ng/mL) | OD | Corrected OD |
---|---|---|
200.00 | 2.328 | 2.247 |
100.00 | 1.556 | 1.475 |
50.00 | 1.067 | 0.986 |
25.00 | 0.895 | 0.814 |
12.50 | 0.587 | 0.506 |
6.25 | 0.265 | 0.184 |
3.13 | 0.189 | 0.108 |
0.00 | 0.081 | 0.000 |
Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
Matrix | Recovery range | Average |
---|---|---|
serum(n=5) | 80-92% | 86% |
EDTA plasma(n=5) | 92-106% | 99% |
Heparin plasma(n=5) | 87-99% | 93% |
Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
---|---|---|---|---|
serum(n=5) | 87-103% | 97-105% | 79-90% | 87-96% |
EDTA plasma(n=5) | 93-99% | 87-101% | 85-95% | 85-96% |
Heparin plasma(n=5) | 96-104% | 90-102% | 87-98% | 95-103% |