Overview

Product name: Mouse CD163(Cluster Of Differentiation) ELISA Kit
Reactivity: Mouse
Alternative Names: M130; MM130; Scavenger receptor cysteine-rich type 1 protein M130; Hemoglobin scavenger receptor
Assay Type: Sandwich
Sensitivity: 13.3 pg/mL
Standard: 2000 pg/mL
Detection Range: 31.25-2000 pg/mL
Sample Type: Serum, plasma and other biological fluids
Assay Length: 3.5h
Research Area: CD & Adhesion molecule;Metabolic pathway;
Uniprot ID: Q2VLH6
Test principle: The test principle applied in this kit is Sandwich enzyme immunoassay. The microtiter plate provided in this kit has been pre-coated with an antibody specific to Mouse CD163. Standards or samples are added to the appropriate microtiter plate wells then with a biotin-conjugated antibody specific to Mouse CD163. Next, Avidin conjugated to Horseradish Peroxidase (HRP) is added to each microplate well and incubated. After TMB substrate solution is added, only those wells that contain Mouse CD163, biotin-conjugated antibody and enzyme-conjugated Avidin will exhibit a change in color. The enzyme-substrate reaction is terminated by the addition of sulphuric acid solution and the color change is measured spectrophotometrically at a wavelength of 450nm ± 10nm. The concentration of Mouse CD163 in the samples is then determined by comparing the OD of the samples to the standard curve.

Standard curve

Concentration (pg/mL) OD Corrected OD
2000.00 2.260 2.179
1000.00 1.527 1.446
500.00 1.142 1.061
250.00 0.872 0.791
125.00 0.509 0.428
62.50 0.319 0.238
31.25 0.288 0.207
0.00 0.081 0.000

Precision

Intra-assay Precision (Precision within an assay):CV%<8%

Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.

Inter-assay Precision (Precision between assays):CV%<10%

Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.

Recovery

Matrices listed below were spiked with certain level of recombinant CD163 and the recovery rates were calculated by comparing the measured value to the expected amount of CD163 in samples.
Matrix Recovery range Average
serum(n=5) 78-92% 85%
EDTA plasma(n=5) 90-105% 97%
Heparin plasma(n=5) 80-97% 88%

Linearity

The linearity of the kit was assayed by testing samples spiked with appropriate concentration of CD163 and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.
Matrix 1:2 1:4 1:8 1:16
serum(n=5) 85-94% 87-101% 86-97% 88-95%
EDTA plasma(n=5) 85-92% 79-96% 92-101% 87-98%
Heparin plasma(n=5) 95-103% 86-92% 89-97% 91-103%