Size: | Price: |
---|---|
48 T | $320.00 |
96 T | $458.00 |
Concentration (ng/mL) | OD | Corrected OD |
---|---|---|
10.00 | 2.073 | 2.000 |
5.00 | 1.628 | 1.555 |
2.50 | 1.032 | 0.959 |
1.25 | 0.782 | 0.709 |
0.63 | 0.543 | 0.470 |
0.32 | 0.335 | 0.262 |
0.16 | 0.168 | 0.095 |
0.00 | 0.073 | 0.000 |
Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
Matrix | Recovery range | Average |
---|---|---|
serum(n=5) | 84-93% | 87% |
EDTA plasma(n=5) | 95-108% | 101% |
Heparin plasma(n=5) | 82-97% | 89% |
Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
---|---|---|---|---|
serum(n=5) | 85-94% | 79-92% | 87-101% | 82-90% |
EDTA plasma(n=5) | 97-105% | 84-96% | 87-98% | 95-102% |
Heparin plasma(n=5) | 79-97% | 90-99% | 95-103% | 89-97% |