Size: | Price: |
---|---|
48 T | $320.00 |
96 T | $458.00 |
Concentration (ng/mL) | OD | Corrected OD |
---|---|---|
100.00 | 2.095 | 1.994 |
50.00 | 1.554 | 1.453 |
25.00 | 1.216 | 1.115 |
12.50 | 0.847 | 0.746 |
6.25 | 0.563 | 0.462 |
3.13 | 0.321 | 0.220 |
1.57 | 0.214 | 0.113 |
0.00 | 0.101 | 0.000 |
Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
Matrix | Recovery range | Average |
---|---|---|
serum(n=5) | 88-99% | 93% |
EDTA plasma(n=5) | 86-98% | 92% |
Heparin plasma(n=5) | 82-98% | 87% |
Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
---|---|---|---|---|
serum(n=5) | 95-104% | 87-96% | 89-101% | 82-90% |
EDTA plasma(n=5) | 95-104% | 85-96% | 87-102% | 91-99% |
Heparin plasma(n=5) | 91-97% | 86-92% | 79-93% | 89-106% |