Size: | Price: |
---|---|
48 T | $320.00 |
96 T | $458.00 |
Concentration (ng/mL) | OD | Corrected OD |
---|---|---|
10.00 | 2.075 | 1.993 |
5.00 | 1.769 | 1.687 |
2.50 | 1.175 | 1.093 |
1.25 | 0.831 | 0.749 |
0.63 | 0.548 | 0.466 |
0.32 | 0.335 | 0.253 |
0.16 | 0.174 | 0.092 |
0.00 | 0.082 | 0.000 |
Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
Matrix | Recovery range | Average |
---|---|---|
serum(n=5) | 83-99% | 95% |
EDTA plasma(n=5) | 85-97% | 91% |
Heparin plasma(n=5) | 85-97% | 91% |
Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
---|---|---|---|---|
serum(n=5) | 87-96% | 85-97% | 85-94% | 87-96% |
EDTA plasma(n=5) | 85-97% | 87-96% | 85-96% | 85-95% |
Heparin plasma(n=5) | 86-96% | 93-104% | 89-97% | 85-94% |