Size: | Price: |
---|---|
48 T | $320.00 |
96 T | $458.00 |
Concentration (ng/mL) | OD | Corrected OD |
---|---|---|
30.00 | 2.157 | 2.058 |
15.00 | 1.527 | 1.428 |
7.50 | 1.196 | 1.097 |
3.75 | 0.839 | 0.740 |
1.88 | 0.527 | 0.428 |
0.94 | 0.236 | 0.137 |
0.47 | 0.187 | 0.088 |
0.00 | 0.099 | 0.000 |
Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
Matrix | Recovery range | Average |
---|---|---|
serum(n=5) | 82-98% | 90% |
EDTA plasma(n=5) | 90-107% | 98% |
Heparin plasma(n=5) | 87-99% | 93% |
Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
---|---|---|---|---|
serum(n=5) | 88-97% | 86-92% | 81-93% | 93-104% |
EDTA plasma(n=5) | 96-105% | 87-96% | 90-98% | 92-101% |
Heparin plasma(n=5) | 93-105% | 79-86% | 87-103% | 85-97% |